(°C)
In contrast to lakes, where DO levels are most likely to vary vertically in the water column, the DO in rivers and streams changes more horizontally along the course of the waterway. This is especially true in smaller, shallower streams. In larger, deeper rivers, some vertical stratification of dissolved oxygen might occur. The DO levels in and below riffle areas, waterfalls, or dam spillways are typically higher than those in pools and slower-moving stretches. If you wanted to measure the effect of a dam, it would be important to sample for DO behind the dam, immediately below the spillway, and upstream of the dam. Since DO levels are critical to fish, a good place to sample is in the pools that fish tend to favor or in the spawning areas they use.
An hourly time profile of DO levels at a sampling site is a valuable set of data because it shows the change in DO levels from the low point just before sunrise to the high point sometime in the midday. However, this might not be practical for a volunteer monitoring program. It is important to note the time of your DO sampling to help judge when in the daily cycle the data were collected.
DO is measured either in milligrams per liter (mg/L) or "percent saturation." Milligrams per liter is the amount of oxygen in a liter of water. Percent saturation is the amount of oxygen in a liter of water relative to the total amount of oxygen that the water can hold at that temperature.
DO samples are collected using a special BOD bottle: a glass bottle with a "turtleneck" and a ground glass stopper. You can fill the bottle directly in the stream if the stream is wadable or boatable, or you can use a sampler that is dropped from a bridge or boat into water deep enough to submerse the sampler. Samplers can be made or purchased. Dissolved oxygen is measured primarily either by using some variation of the Winkler method or by using a meter and probe.
Winkler Method
The Winkler method involves filling a sample bottle completely with water (no air is left to bias the test). The dissolved oxygen is then "fixed" using a series of reagents that form an acid compound that is titrated. Titration involves the drop-by-drop addition of a reagent that neutralizes the acid compound and causes a change in the color of the solution. The point at which the color changes is the "endpoint" and is equivalent to the amount of oxygen dissolved in the sample. The sample is usually fixed and titrated in the field at the sample site. It is possible, however, to prepare the sample in the field and deliver it to a lab for titration.
Dissolved oxygen field kits using the Winkler method are relatively inexpensive, especially compared to a meter and probe. Field kits run between $35 and $200, and each kit comes with enough reagents to run 50 to 100 DO tests. Replacement reagents are inexpensive, and you can buy them already measured out for each test in plastic pillows.
You can also buy the reagents in larger quantities, in bottles, and measure them out with a volumetric scoop. The advantage of the pillows is that they have a longer shelf life and are much less prone to contamination or spillage. The advantage of buying larger quantities in bottles is that the cost per test is considerably less.
The major factor in the expense of the kits is the method of titration they use eyedropper, syringe-type titrator, or digital titrator. Eyedropper and syringe-type titration is less precise than digital titration because a larger drop of titrant is allowed to pass through the dropper opening and, on a micro-scale, the drop size (and thus the volume of titrant) can vary from drop to drop. A digital titrator or a buret (which is a long glass tube with a tapered tip like a pipet) permits much more precision and uniformity in the amount of titrant that is allowed to pass.
If your program requires a high degree of accuracy and precision in DO results, use a digital titrator. A kit that uses an eye dropper-type or syringe- type titrator is suitable for most other purposes. The lower cost of this type of DO field kit might be attractive if you are relying on several teams of volunteers to sample multiple sites at the same time.
Meter and Probe
A dissolved oxygen meter is an electronic device that converts signals from a probe that is placed in the water into units of DO in milligrams per liter. Most meters and probes also measure temperature. The probe is filled with a salt solution and has a selectively permeable membrane that allows DO to pass from the stream water into the salt solution. The DO that has diffused into the salt solution changes the electric potential of the salt solution and this change is sent by electric cable to the meter, which converts the signal to milligrams per liter on a scale that the volunteer can read.
DO meters are expensive compared to field kits that use the titration method. Meter/probe combinations run between $500 and $1,200, including a long cable to connect the probe to the meter. The advantage of a meter/probe is that you can measure DO and temperature quickly at any point in the stream that you can reach with the probe. You can also measure the DO levels at a certain point on a continuous basis. The results are read directly as milligrams per liter, unlike the titration methods, in which the final titration result might have to be converted by an equation to milligrams per liter.
However, DO meters are more fragile than field kits, and repairs to a damaged meter can be costly. The meter/probe must be carefully maintained, and it must be calibrated before each sample run and, if you are doing many tests, in between samplings. Because of the expense, a volunteer program might have only one meter/probe. This means that only one team of samplers can sample DO and they will have to do all the sites. With field kits, on the other hand, several teams can sample simultaneously.
Laboratory Testing of Dissolved Oxygen
If you use a meter and probe, you must do the testing in the field; dissolved oxygen levels in a sample bottle change quickly due to the decomposition of organic material by microorganisms or the production of oxygen by algae and other plants in the sample. This will lower your DO reading. If you are using a variation of the Winkler method, it is possible to "fix" the sample in the field and then deliver it to a lab for titration. This might be preferable if you are sampling under adverse conditions or if you want to reduce the time spent collecting samples. It is also a little easier to titrate samples in the lab, and more quality control is possible because the same person can do all the titrations.
The procedures for collecting and analyzing samples for dissolved oxygen consist of the following tasks:
Refer to section 2.3 - Safety Considerations for details on confirming sampling date and time, safety considerations, checking supplies, and checking weather and directions. In addition to the standard sampling equipment and apparel, when sampling for dissolved oxygen, include the following equipment:
If Using the Winkler Method
If Using a Meter and Probe
The directions for sampling should provide specific information about the exact point in the stream from which you are to sample; e.g., "approximately 6 feet out from the large boulder downstream from the west side of the bridge." If you are not sure you are in the exact spot, record a detailed description of where you took the sample so that it can be compared to the actual site later.
Use a BOD bottle to collect the water sample. The most common sizes are 300 milliliters (mL) and 60 mL. Be sure that you are using the correct volume for the titration method that will be used to determine the amount of DO. There is usually a white label area on the bottle, and this may already be numbered. If so, be sure to record that number on the field data sheet. If your bottle is not already numbered, place a label on the bottle (not on the cap because a cap can be inadvertently placed on a different bottle) and use a waterproof marker to write in the site number.
If you are collecting duplicate samples, label the duplicate bottle with the correct code, which should be determined prior to sampling by the lab supplying the bottles. Use the following procedure for collecting a sample for titration by the Winkler method:
Point the bottle downstream and fill gradually. Cap underwater when full. |
Using a DO Meter
If you are using a dissolved oxygen meter, be sure that it is calibrated immediately prior to use. Check the cable connection between the probe and the meter. Make sure that the probe is filled with electrolyte solution, that the membrane has no wrinkles, and that there are no bubbles trapped on the face of the membrane. You can do a field check of the meter's accuracy by calibrating it in saturated air according to th e manufacturer's instructions. Or, you can measure a water sample that is saturated with oxygen, as follows. (NOTE: You can also use this procedure for testing the accuracy of the Winkler method.)
Once the meter is turned on, allow 15 minute equilibration before calibrating. After calibration, do not turn the meter off until the sample is analyzed. Once you have verified that the meter is working properly, you are ready to measure the DO levels at the sampling site. You might need an extension pole (this can be as simple as a piece of wood) to get the probe to the proper sampling point. Simply secure the probe to the end of the extension pole. A golfer's ball retriever works well because it is collapsible and easy to transport. To use the probe, proceed as follows:
Three types of titration apparatus can be used with the Winkler method: droppers, digital titrators, and burets. The dropper and digital titrator are suited for field use. The buret is more conveniently used in the lab (Fig. 5.8) Volunteer programs are most likely to use the dropper or digital titrator. For titration with a dropper or syringe, which is relatively simple, follow the manufacturer's instructions. The following procedure is for using a digital titrator to determine the quantity of dissolved oxygen in a fixed sample:
|
Some water quality standards are expressed in terms of percent saturation. To calculate percent saturation of the sample:
Expected Range | Sample Volume | Titration Cartridge | Digit Multiplier | |
1-5 mg/L | 200 mL | 0.2 N | 0.01 | |
2-10 mg/L | 100 mL | 0.2 N | 0.02 | |
10+ mg/L | 200 mL | 2.0 N | 0.10 |
If you are using the Winkler method and delivering the samples to a lab for titration, double-check to make sure that you have recorded the necessary information for each site on the field data sheet, especially the bottle number and corresponding site nu mber and the times the samples were collected. Deliver your samples and field data sheets to the lab. If you have already obtained the dissolved oxygen results in the field, send the data sheets to your sampling coordinator.
Biochemical oxygen demand, or BOD, measures the amount of oxygen consumed by microorganisms in decomposing organic matter in stream water. BOD also measures the chemical oxidation of inorganic matter (i.e., the extraction of oxygen from water via chemical reaction). A test is used to measure the amount of oxygen consumed by these organisms during a specified period of time (usually 5 days at 20 C). The rate of oxygen consumption in a stream is affected by a number of variables: temperature, pH, the presence of certain kinds of microorganisms, and the type of organic and inorganic material in the water.
BOD directly affects the amount of dissolved oxygen in rivers and streams. The greater the BOD, the more rapidly oxygen is depleted in the stream. This means less oxygen is available to higher forms of aquatic life. The consequences of high BOD are the same as those for low dissolved oxygen: aquatic organisms become stressed, suffocate, and die.
Sources of BOD include leaves and woody debris; dead plants and animals; animal manure; effluents from pulp and paper mills, wastewater treatment plants, feedlots, and food-processing plants; failing septic systems; and urban stormwater runoff.
BOD is affected by the same factors that affect dissolved oxygen (see above). Aeration of stream water by rapids and waterfalls, for example will accelerate the decomposition of organic and inorganic material. Therefore, BOD levels at a sampling site with slower, deeper waters might be higher for a given volume of organic and inorganic material than the levels for a similar site in highly aerated waters.
Chlorine can also affect BOD measurement by inhibiting or killing the microorganisms that decompose the organic and inorganic matter in a sample. If you are sampling in chlorinated waters, such as those below the effluent from a sewage treatment plant, it is necessary to neutralize the chlorine with sodium thiosulfate. (See APHA, 1992.)
BOD measurement requires taking two samples at each site. One is tested immediately for dissolved oxygen, and the second is incubated in the dark at 20 C for 5 days and then tested for the amount of dissolved oxygen remaining. The difference in oxygen levels between the first test and the second test, in milligrams per liter (mg/L), is the amount of BOD. This represents the amount of oxygen consumed by microorganisms to break down the organic matter present in the sample bottle during the incubation period. Because of the 5-day incubation, the tests should be conducted in a laboratory.
Sometimes by the end of the 5-day incubation period the dissolved oxygen level is zero. This is especially true for rivers and streams with a lot of organic pollution. Since it is not known when the zero point was reached, it is not possible to tell what the BOD level is. In this case it is necessary to dilute the original sample by a factor that results in a final dissolved oxygen level of at least 2 mg/L. Special dilution water should be used for the dilutions. (See APHA, 1992.)
It takes some experimentation to determine the appropriate dilution factor for a particular sampling site. The final result is the difference in dissolved oxygen between the first measurement and the second after multiplying the second result by the dilution factor. More details are provided in the following section.
The procedures for collecting samples for BOD testing consist of the same steps described for sampling for dissolved oxygen (see above), with one important difference. At each site a second sample is collected in a BOD bottle and delivered to the lab for DO testing after the 5-day incubation period. Follow the same steps used for measuring dissolved oxygen with these additional considerations:
The first bottle should be analyzed just prior to storing the second sample bottle in the dark for 5 days at 20 C. After this time, the second bottle is tested for dissolved oxygen using the same method that was used for the first bottle. The BOD i s expressed in milligrams per liter of DO using the following equation:
DO (mg/L) of first bottle - DO (mg/L) of second bottle = BOD (mg/L)
APHA. 1992. Standard methods for the examination of water and wastewater. 18 th ed. American Public Health Association, Washington, DC.
Last updated on Tuesday, March 06, 2012
The Seal of the United States Environmental Protection Agency
Jump to main content.
View Issue TOC Volume 34, No. 2 Pages 177 - 183
By email: [email protected] affiliation: dauphin island sea lab, dauphin island, al, usa, and university of south alabama, mobile, al, usa search for more papers by this author ">kara j. gadeken and affiliation: dauphin island sea lab, dauphin island, al, usa, and university of south alabama, mobile, al, usa search for more papers by this author ">kelly m. dorgan .
Dauphin Island Sea Lab, Dauphin Island, AL, USA, and University of South Alabama, Mobile, AL, USA
Purpose of device.
Changes in dissolved oxygen concentration can cause dramatic shifts in chemical, biological, and ecological processes in aquatic systems. In shallow coastal areas, this can happen on short timescales, with oxygen increasing during the day due to photosynthesis and declining at night due to respiration. We present a system controlled by an Arduino microprocessor that leverages the oxygen-consuming capacity of sediments to manipulate dissolved oxygen in an aquarium tank to planned concentrations. With minor adjustments to the Arduino code, the system can produce a variety of dissolved oxygen patterns, including a diel cycle. Designed to be user-friendly and scalable if needed, the system uses easily acquired, low-cost electronic and aquarium components. Its simplicity and accessibility permit deeper exploration of the effects of dissolved oxygen variability in aquatic systems, and the use of Arduino code and basic electronics makes it a potential tool for teaching experimental design and instrument fabrication.
The availability of dissolved oxygen (DO) is a major factor governing aquatic ecosystem function and is an indicator of water quality and ecosystem health (Diaz and Rosenberg, 1995, 2011; Wenner et al., 2004; Middelburg and Levin, 2009). The DO concentration in aquatic environments is controlled by the balance of oxygen sources (mixing with the atmosphere, advection of oxygenated water, photosynthetic production) and sinks (aerobic respiration and abiotic oxidation), and shifts in this balance result in cascading chemical, biological, and ecological effects (Middelburg and Levin, 2009). Changes in DO concentration occur across temporal and spatial scales, from widespread, seasons-long bottom hypoxia on continental margins to dramatic daily or sometimes hourly oxygen fluctuations in shallow, semi-enclosed coastal lagoons or embayments. Most low oxygen events are this second type, relatively short in duration but occurring frequently (Wenner et al., 2004). Many researchers have examined the effects of declining or persistent low DO on water and sediment chemistry (McCarthy et al., 2008; Lehrter et al., 2012; Neubacher et al., 2013; Foster and Fulweiler, 2019) and organismal behavior and physiology (Diaz and Rosenberg, 1995; Long et al., 2008; Levin et al., 2009; Sturdivant et al., 2012; Riedel et al., 2014; Calder-Potts et al., 2015). However, it is far less common to see investigations into dynamic variation in DO, likely because of the difficulties in precisely and repeatedly manipulating DO in the lab.
DO can easily be increased in water by bubbling with air, but decreasing DO requires either chemical consumption or physical expulsion of oxygen from solution. An often-used method of decreasing DO involves stripping it from the water by bubbling with N 2 gas. Studies of low oxygen effects that run for multiple weeks or months, however, may require large amounts of N 2 gas, which can be expensive, prompting investigations into ways to reduce the amount of gas needed (Bevan and Kramer, 1988; Peterson and Ardahl, 1992; Grecay and Stierhoff, 2002). Oxygen can also be removed by “vacuum degassing,” applying a partial vacuum to the water to remove DO from solution. This requires an airtight vacuum setup that may not be feasible for some researchers (Mount, 1961; Miller et al., 1994). More recently, researchers have developed methods that rely on chemical consumption rather than physical removal of DO to produce low oxygen water. Thetmeyer et al. (1999) leveraged the respiration of the fish study subject itself to draw down DO, controlling hypoxic, normoxic, and oscillating oxygen treatments with an automated system (Thetmeyer et al., 1999). For this method to work, the fish must consume enough oxygen to change the DO of the experimental environment, which may not be possible for smaller study subjects or those for which wall effects are a concern. Long et al. (2008) presented an alternative method using sediments to decrease DO by percolating water through a “fluidized mud reactor” that consumed oxygen (Long et al., 2008). The resulting anoxic water was then mixed in different proportions with fully oxygenated water to produce predetermined DO concentrations. This setup is convenient for creating water with stable DO concentrations, but it does not easily allow for complex manipulations of DO change through time.
The existing DO manipulation methods pose a barrier to entry for many researchers because of their costs and complexities. Additionally, many methods have been designed to simulate long-term hypoxia, whereas in coastal systems, DO concentrations can vary on short timescales. Here, we describe and test a DO manipulation system that can be constructed in a laboratory or classroom setting using easily acquired electronic and aquarium components. The closed-loop system does not require N 2 gas purging or vacuum degassing; instead, it relies on sediment oxygen demand (SOD) to draw DO down, and it increases DO by periodically opening a solenoid valve to allow oxygenated water to flow in from an upstream reservoir tank ( Figure 1 ). The provided code is uploaded to an Arduino microprocessor that monitors and adjusts the DO in the experiment tank to a pattern planned by the user, simultaneously recording and displaying the DO data. This system was built to study behavior of and SOD by infaunal organisms held individually in small sediment-filled aquaria ( Figure 1 ), but it could be used for a variety of shallow-water systems and study organisms. This simple, low-cost, open-source method of manipulating DO in the lab will allow for more varied studies into how change in DO affects aquatic systems.
Diagram of oxygen manipulation setup. It is a closed-loop system, constantly cycling water between an oxygenated reservoir tank and a sump. When the dissolved oxygen (DO) in the experiment tank is sensed to be lower than the desired level due to sediment consumption, the solenoid valve is opened, allowing oxygenated water to flow in. A power head in the experiment tank ensures that the water is well mixed, and a layer of bubble wrap floating at the water surface prevents diffusion of atmospheric oxygen into the water. PVC pipe is shown in white, tubing in gray, and wiring in black. Note that, though only one experiment tank is depicted here, several replicate experiment tanks would be needed or several replicate trials should be performed to avoid pseudoreplication. |
Table 1 outlines the system components. We estimate the total cost of constructing this system to be approximately US$625 in 2021, not including shipping expenses. Note that the components list consists only of consumable items (e.g., wire, tubing, plumbing) and specialized equipment (e.g., Arduino, Atlas Scientific EZO DO kit, pumps, solenoid valve). Non-consumables and tools needed for assembly are not included because it is assumed that the user will already have access to many of these items. Cost of construction might also be substantially lowered if materials can be purchased individually rather than in large packs, or scavenged from other projects, as only a single item or a very small amount is required for most components. An undergraduate or a particularly capable high school student should be able to construct and begin using the system within a couple of weeks.
Components list for laboratory dissolved oxygen experiments. |
First, assemble the electronics according to the wiring diagram in Figure 2 . The Arduino Mega®️, solenoid, and relay may all be powered by the same 12V power source. The Adafruit® data logging shield is mounted directly to the Arduino via soldered header pins. The Atlas Scientific®️ EZO DO circuit is mounted on an electrically isolated EZO carrier board and connects to the SCL (clocking) and SDA (data-transmitting) pins on the data logging shield to communicate with the Arduino ( Figure 2 ). Communication between the Arduino and the EZO DO circuit is in I 2 C protocol to allow for easy addition of secondary devices, in case more circuits and sensors are desired to scale up the system. The EZO DO circuit must be converted to I 2 C protocol and the I 2 C address changed to correspond to the address defined in the oxygen manipulation code to communicate with the Arduino. The EZO DO circuit should also be adjusted with temperature and salinity offsets and two-point oxygen calibrated before each use. Directions on conversion to I 2 C protocol, offset adjustments, and calibration are in the EZO DO circuit documentation. An LCD screen is included to display the average measured DO over the previous several measurements and the planned DO, allowing the user to easily assess whether the system is functioning properly and following the prescribed pattern. Power to the LCD can be converted from 12V to 5V DC with a power converter, as shown in Figure 2 , or sourced from the 5V pin on the Arduino. A master on/off power switch is also included and a small push button is wired in to control when the oxygen manipulation code begins (“start” button).
Wiring diagram. AC power from the grid is converted into DC power, shown as red (VCC) and black (GND) wiring. The VCC terminal block distributes power to each component, and the GND terminal block is a common ground to close the circuit. The Atlas Scientific EZO DO circuit and the LCD screen are controlled via I C protocol from the SCL (clocking) and SDA (data) pins. The Arduino Mega 2560, SD shield, and LCD screen images are from , and the Atlas Scientific EZO DO circuit image is from circuit documentation on . |
When the code is started, it executes in repeated “loops”; within a single loop, the system measures the DO in the tank, calculates the average DO over the last five readings, compares the average DO to the programmed DO, opens the valve to allow in oxygenated water if necessary, logs the data to the SD card, and displays the average and planned DO on the LCD screen. The code may be restarted by pressing the “reset” button on the SD shield and then the wired-in “start” button. The average DO value is used rather than the instantaneous DO value to adjust for inherent noise in measurements (i.e., to prevent the system from allowing a single anomalously low measurement to trigger oxygenated water to flow in, even when the average DO is above the planned level).
Once the electronics have been assembled, construct the closed loop tank system ( Figure 1 ). Three tanks are used: the upstream reservoir tank for oxygenated water, the experiment tank in which study subjects are held and DO is manipulated, and a sump. Oxygenated water will constantly circulate between the reservoir and the sump, and it will be intermittently diverted into the experiment tank whenever DO needs to be increased.
The reservoir tank and the sump may be made from simple plastic bins. Ideally, the experiment tank should be a clear aquarium tank so study subjects may be easily observed. Fit the experiment tank with an outflow standpipe and fill the tank with a layer of organic-rich sediment, which will consume oxygen and drive down DO in the overlying water. Fill the remaining space in the tank up to the top of the standpipe with seawater and allow suspended sediment to settle. Then, make two outflow holes in the reservoir tank. Attach a standpipe to the first hole and add plumbing to the outlet to direct overflow water into the sump. Attach the solenoid valve to the second hole and add piping or tubing to direct flow from the solenoid into the experiment tank. Position the reservoir tank higher than the experiment tank and fill the reservoir tank with seawater up to the standpipe. Place a sump pump in the sump and route tubing from its outflow up to the reservoir tank to close the loop.
Suspend the Atlas Scientific DO sensor in the experiment tank. The water’s surface should be covered by a sheet of bubble wrap (which is oxygen-impermeable and will float at the surface) to prevent diffusion of atmospheric oxygen into the water. A small aquarium power head should be mounted in the experiment tank to gently circulate the water and prevent stagnation. In our setup with a 20 gallon experiment tank, a 120 gph power head was sufficient. Manipulations by the Arduino are based on the readings from the sensor, so it is critical that the water be mixed such that the sensor readings represent the DO of the bulk water in the tank as accurately as possible. It is also important to note that if the system is to be used for rigorous experimental work, having all replicates in one tank presents the issue of pseudoreplication. To resolve this issue, multiple replicate experiment tanks should be plumbed and manipulated “in parallel,” or if it is only feasible to have one experiment tank, multiple replicate trials should be performed over time.
The annotated oxygen manipulation code and the calibration code are freely accessible for download on the code sharing platform GitHub ( https://github.com/kgadeken/OxygenManipulationCode_GadekenDorgan2021 ). It is highly recommended that new users read the code and annotations thoroughly before setting up and using the system.
To serve as a usable method for manipulating oxygen in the lab, the system must reliably, precisely, and accurately produce the programmed DO patterns in the experiment tank. We tested the system by programming it to generate a diel oxygen cycle, with DO concentrations ranging from 3 mg L –1 to 7 mg L –1 . The diel cycle spans a wide range of DO values and demands the system adapt quickly to continually varying rates of DO change, making it a highly rigorous test of the system’s flexibility. Precision was gauged by the difference of each DO measurement from the programmed DO value at that time. To gauge the system’s accuracy, we took corroborating oxygen measurements with an Onset HOBO DO logger. The Atlas Scientific probe and HOBO logger were both two-point calibrated immediately before starting the trial. The Atlas Scientific probe and the HOBO logger were secured in the experiment tank as close together as possible in the upper-middle of the water column at the same vertical height from the sediment surface. The HOBO logger was set to measure DO every five minutes.
Figure 3 shows the results of the diel cycle trial. The system closely followed the diel pattern during rising and high DO periods but deviated slightly during falling DO. This indicates that the sediment was not consuming enough oxygen at these times to keep up with the programmed rate of decrease. At its greatest point, the difference between the measured DO and the planned DO was 0.91 mg L –1 . However, over 99% of the measurements taken deviated from the planned value by less than half of that maximum difference (±0.46 mg L –1 ), and ~89% deviated by less than a quarter (±0.23 mg L –1 ), indicating that the system typically followed the programmed pattern very closely.
Results of testing of the oxygen manipulation system using a diel oxygen cycle (ranging from 3 mg L to 7 mg L ). Precision was assessed from the difference between the measured DO (blue) and the programmed DO (gray) at each time point. An Onset HOBO DO logger (yellow) was included to take corroborating measurements every five minutes to assess the accuracy of the system’s oxygen measurements and manipulation. |
Because the HOBO was set to take measurements at 5 min intervals while the oxygen manipulation system measured DO every 37 s, values were interpolated from the HOBO sensor measurements to correspond to each of the measurements from the oxygen manipulation system. The HOBO measurements followed the same diel pattern as the system but were positively offset an average of 0.62 mg L –1 . The data from the oxygen manipulation system and the HOBO were both detrended and then analyzed for correlation, and no lag was found between the two data sets. Given this result, it is likely that the difference between the sensor measurements is largely due to calibration error.
A potential issue with this setup lies in the use of sediment oxygen demand as a DO sink. Using SOD rather than N 2 gas limits the rate of oxygen removal from the water, as shown in the diel cycle trial during the periods of DO decline ( Figure 3 ). Vigorous purging with N 2 gas or vacuum degassing can remove all oxygen from solution in seconds or minutes (Mount, 1961), whereas our system using SOD typically takes several hours to decrease from full oxygen saturation to hypoxia. Also, low DO is well known to be accompanied by a suite of related changes in sediment and water chemistry, including altered chemical concentrations, changes in nutrient flux, and modified pH (Froelich et al., 1979; Burnett, 1997; Middelburg and Levin, 2009). This is in contrast to the N 2 gas and vacuum degassing methods that strip DO by physically removing it from the water, and thus do not result in the same chemical reactions as SOD. However, using sediments to scrub oxygen more closely resembles how low DO occurs in situ. Bubbling water with N 2 gas to remove oxygen decreases the p CO 2 of the water, and so increases pH, in contrast to oxygen consumption by sediments that typically decreases pH because organic matter remineralization generates CO 2 (Gobler et al., 2014). Though oxygen cannot be drawn down as quickly and the effects of change in DO alone cannot be as cleanly isolated with the SOD method because other chemical characteristics are unavoidably covarying, it more accurately represents DO variability as it would be encountered in natural settings. Also, because this system is closed loop and relies on biological processes to function, issues with excessive buildup of ammonia and nitrates may arise if experiments are run for extended periods without replacing the water in the tanks. This system is best applied in situations that do not require independent control of water chemistry variables and for experiments that can be performed within a short time frame.
There are several ways that the user can modify the system to work more effectively or to troubleshoot issues. We divide them into “out-code” modifications, or changes to certain physical or structural features in the system, and “in-code” modifications, or changes to the Arduino code that alter the way the oxygen manipulation is executed.
The efficacy of the system depends heavily on the oxygen-consuming capacity of the sediments in the experiment tank. Use the most organic-rich sediments available and maximize the ratio of sediment surface area to bulk water volume by using as shallow a tank as possible. Before starting construction of the system, we recommend assessing the oxygen consumption rate the mud can achieve by putting mud into a tank with overlying water to the height anticipated for the experiment, adding a power head to circulate the water, covering the water with bubble wrap, and recording the oxygen through time. If the sediment is not consuming oxygen at a sufficient rate, adding labile organic matter or fertilizer to the sediment or displacing some of the overlying water with a solid object can help increase oxygen drawdown.
Although the power head in the experiment tank may be effective at laterally circulating water, there is still potential for vertical DO gradients to form in the tank, and the steepness of the gradient will increase closer to the sediment surface. Thus, the positioning of the probe in the tank is important. The probe should be secured in position well above the sediment surface and close to the vertical level where study organisms will likely be located. Because the code manipulates DO based on the readings from only one probe, it is also critical to take care in calibrating the sensor, as exemplified by the diel cycle trial ( Figure 3 ). Before it is used, the sensor should be two-point calibrated and its accuracy corroborated at multiple DO concentrations with measurements from a reliable instrument. Though the calibration held well in the diel cycle trial, during longer experiments we advise periodically comparing the DO concentration against a reliable measurement to check for sensor drift.
Because this system was devised for a study designed to observe responses of sediment-dwelling organisms to changing DO, the study organisms were kept in smaller replicate containers filled with sediment within the experiment tank. We constructed a platform that sat on stilts just above the sediment layer to support the replicate containers ( Figure 1 ). This platform had as many gaps as possible so that the water at the sediment surface was well mixed.
Two main features of the code may easily be altered to change the way that DO control is performed: the amount of time the solenoid is held open, and the pause duration between loops.
If DO data are noisy and repeatedly jump substantially above the planned DO concentration, too much oxygenated water may be flowing in with each loop, and the amount of time the solenoid is held open should be decreased. Decreasing the duration between each loop changes the frequency with which the DO is compared to the planned value and manipulated, essentially changing the sensitivity of the system. If this duration is too short, the power head in the tank may not have enough time to circulate the high oxygen water added in the previous loop, resulting in inaccurate measurements and manipulation. The time the solenoid is held open and the pause duration between loops work in concert to affect the precision of DO manipulation, and some amount of trial and error will be necessary to determine the optimal values for each. That said, the system has proven to be somewhat resilient to changes in these variables. We performed a sensitivity test by programming it to maintain DO at 5 mg L –1 for ~1.5 h four times, each with a different combination of the amount of time the valve is left open (either 3 s or 6 s) and the time between loops (either 20 s or 1 min) ( Figure 4 ). In the four trials (3s:20s, 3s:1min, 6s:20s, and 6s:1min), the maximum deviation of the measured DO from the planned DO was 0.21, 0.27, 0.25, and 0.23 mg L –1 , respectively. The percentage of time that the measured value deviated by less than half the maximum deviation was 80%, 91%, 82%, and 81%, respectively. All four trials effectively maintained the programmed DO concentration within a small range of variability.
Results from sensitivity testing. The system was run four times, programmed to maintain DO at 5 mg L (gray line) for ~1.5 h using different combinations of the amount of time the valve was held open, either 3 s (a and b) or 6 s (c and d), and the time waited between loops, either 20 s (a and c) or 1 min (b and d). |
Aquatic organisms, particularly in coastal areas, exist in an environment with complex variations in DO that have long been difficult to reproduce in a lab setting. Perhaps the most compelling prospect of the described system is its capacity to replicate this variation for study. The programmed oxygen pattern is controlled directly through the Arduino code to allow greater flexibility in the choice, combination, and order of programmed patterns. For example, simply by adjusting the value or equation that the Arduino code is programmed to match and re-uploading the code, the system could be made to alternate increasing and decreasing DO at specific rates, allowing more rigorous study of how the rate of increase or decrease in oxygen affects animal behavior. Or, as was shown in the system test, it can produce a pattern from a modified sine function that mimics a diel oxygen cycle, an extremely common oxygen pattern in shallow coastal waters that remains understudied. The system could further be retrofitted with a high-pressure valve and a small N 2 tank for supplementing with N 2 purging when a more rapid oxygen decrease is needed, or the oxygen probe could be upgraded to an optical sensor for more accurate and precise oxygen manipulation.
The system is also potentially useful for educational applications. It is designed to be as simple and modular as possible, with readily available and reasonably priced components and relatively easy construction. Furthermore, using the code requires some familiarity with the Arduino programming language and can serve as a model of how to use Arduino to build instrumentation for scientific inquiry. The system could be equally employed for classroom behavioral or physiological experiments and as a tool to teach experimental instrument fabrication.
This project was made possible by grants from the gulf of mexico research initiative (gomri) through the alabama center for ecological resilience (acer) consortium and the national science foundation (nsf-oce 1844910 to kmd). kg was supported by a graduate fellowship from the department of marine sciences, university of south alabama. thanks to behzad mortazavi and grant lockridge for helpful discussions..
Gadeken, K.J., and K.M. Dorgan. 2021. A simple and inexpensive method for manipulating dissolved oxygen in the lab. Oceanography 34(2):177–183, https://doi.org/10.5670/oceanog.2021.202 .
This is an open access article made available under the terms of the Creative Commons Attribution 4.0 International License ( https://creativecommons.org/licenses/by/4.0/ ), which permits use, sharing, adaptation, distribution, and reproduction in any medium or format as long as users cite the materials appropriately, provide a link to the Creative Commons license, and indicate the changes that were made to the original content. Images, animations, videos, or other third-party material used in articles are included in the Creative Commons license unless indicated otherwise in a credit line to the material. If the material is not included in the article’s Creative Commons license, users will need to obtain permission directly from the license holder to reproduce the material.
Remember Me
Dissolved oxygen in water.
Experiment #12A from Advanced Biology with Vernier
Although water is composed of oxygen and hydrogen atoms, biological life in water depends upon another form of oxygen—molecular oxygen. Oxygen is used by organisms in aerobic respiration, where energy is released by the combustion of sugar in the mitochondria. This form of oxygen can fit into the spaces between water molecules and is available to aquatic organisms.
Fish, invertebrates, and other aquatic animals depend upon the oxygen dissolved in water. Without this oxygen, they would suffocate. Some organisms, such as salmon, mayflies, and trout, require high concentrations of oxygen in their water. Other organisms, such as catfish, midge fly larvae, and carp can survive with much less oxygen. The ecological quality of the water depends largely upon the amount of oxygen the water can hold. The quality of the water can be assessed with fair accuracy by observing the aquatic animal populations in a stream.
In this experiment, you will
This experiment features the following sensors and equipment. Additional equipment may be required.
Ask an expert.
Get answers to your questions about how to teach this experiment with our support team.
This experiment is #12A of Advanced Biology with Vernier . The experiment in the book includes student instructions as well as instructor information for set up, helpful hints, and sample graphs and data.
Pasco partnerships.
What factors affect dissolved oxygen.
Students will use a dissolved oxygen sensor to determine how different factors affect the concentration of dissolved oxygen in water.
Grade Level: High School
Subject: Biology • Environmental Science
Student Files
123.42 KB | ||
125.54 KB |
Teacher Files
Sign In to your PASCO account to access teacher files and sample data.
This sensor is perfect for the live or long-term monitoring of dissolved oxygen concentration and saturation, in both the lab and field.
Many lab activities can be conducted with our Wireless , PASPORT , or even ScienceWorkshop sensors and equipment. For assistance with substituting compatible instruments, contact PASCO Technical Support . We're here to help. Copyright © 2019 PASCO
High School
Middle School
IMAGES
VIDEO
COMMENTS
Laboratory Experiment 6: Dissolved Oxygen (DO) Objective: Determine DO content of a given sample Background: Dissolved oxygen (DO) levels in environmental water depend on the physiochemical and biochemical activities in water body and it is an important useful in pollution and waste treatment process control. ...
The dissolved oxygen concentration of seawater is defined as the number of milliliters of dioxygen gas (O2 ) per liter of seawater (mL L -1 ). 4. Principle of Analysis. The chemical determination of oxygen concentrations in seawater is based on the method first proposed by Winkler (1888) and modified by Strickland and Parsons (1968).
The Winkler Method is a technique used to measure dissolved oxygen in freshwater systems. Dissolved oxygen is used as an indicator of the health of a water body, where higher dissolved oxygen concentrations are correlated with high productivity and little pollution. This test is performed on-site, as delays between sample collection and testing ...
The Winkler method for dissolved oxygen (DO) is a standard titration technique to measure the oxygen content in water. Water is collected in a sample bottle with no air that could interfere with the DO reading. Fixed reagents are added to the sample to form an acidic compound. The sample is then neutralized using a neutralizing compound to ...
1.1 Dissolved Oxygen (DO) Standardization of Na 2 S2 O 3 Solution Potassium bi-iodate (KH(IO 3)2) is first reduced to an equivalent amount of iodine by the addition of KI. This iodine is then quantified by titration with Na 2 S2 O 3 as described in experiment 5A. Note: Write the relevant equations for the above reactions. Procedure:
Dear viewer/subscriber, if my videos helped you a lot (maybe you aced your exams as a student, or you won the admiration and full attention of your students ...
Dissolved oxygen (DO) is a measure of how much oxygen is dissolved in a sample. The amount of dissolved oxygen in water can tell us about the water quality. ...
Saturation of 91 to 110% dissolved oxygen is considered excellent; between 71 and 90% is good, 51-70% is fair, and below 50% is poor. Dissolved oxygen levels of 6 mg/L are sufficient to support most aquatic species. Levels below 4 mg/L are stressful to the majority of aquatic animals, so biodiversity will be affected.
Introduction. Dissolved oxygen is one of the primary indicators of the quality of an aquatic environment. Oxygen enters water from the surrounding air, as a product of photosynthesis, and as a result of rapid movement of water. A Dissolved Oxygen Probe can be used in a wide variety of tests or investigations to determine dissolved oxygen ...
Experiment 5: Dissolved Oxygen (4500-O. C. Azide Modification) OBJECTIVES: To determine Dissolved Oxygen (D.O) concentration in a water sample. BACKGROUND AND PRINCIPLE: The test was first developed by Lajos Winkler while working on his doctoral dissertation in 1888. Dissolved oxygen (D.O) levels in environmental water depend on the physiochemical
Dissolved oxygen levels can be measured by a basic chemical analysis method (titration method), an electrochemical analysis method (diaphragm electrode method), and a photochemical analysis method (fluorescence method). The diaphragm electrode method is the most widely used method. Titration Method. Diaphragm Electrode Method.
06) Aquatic Photosynthesis and Dissolved Gases. Students use a Wireless Optical Dissolved Oxygen sensor (PS-3246), Wireless CO₂ sensor (PS-3208), Dissolved CO₂ Waterproof Sleeve (PS-3545), and a Photosynthesis Chamber (PS-3251) to compare the metabolic products of Elodea in light and dark conditions.
Testing dissolved oxygen (DO) in water is either measured via chemical analysis such as a titrimetric method, electroanalytical (using galvanic & ... Mini Lab Grade Dissolved Oxygen Probe $ 134.99 USD. Add to cart. Lab Grade Dissolved Oxygen Probe $ 259.99 - $ 261.99 USD. Select options. Industrial Dissolved Oxygen Probe $ 349.99 USD.
Dissolved Oxygen and Temperature Lab Background: In this experiment, you will try to find out what happens to the amount of dissolved oxygen when the temperature of the water increases or decreases. This will help you understand how seasonal changes as well as long-term changes, such as climate change, might affect aquatic ecosystems.
Field and lab meters to measure dissolved oxygen have been around for a long time. As this picture shows, modern meters are small and highly electronic. They still use a probe, which is located at the end of the cable. Dissolved oxygen is dependent on temperature (an inverse relation), so the meter must be calibrated properly before each use.
1.0. Oxygen gas is dissolved in water by a variety of processes—diffusion between the atmosphere and water at its surface, aeration as water flows over rocks and other debris, churning of water by waves and wind, and photosynthesis of aquatic plants. There are many factors that affect the concentration of dissolved oxygen in an aquatic ...
Many lab activities can be conducted with our Wireless, PASPORT, or even ScienceWorkshop sensors and equipment. For assistance with substituting compatible instruments, contact PASCO Technical Support. We're here to help. Students will use a dissolved oxygen sensor to determine how different factors affect the concentration of dissolved oxygen ...
Laboratory Testing of Dissolved Oxygen. If you use a meter and probe, you must do the testing in the field; dissolved oxygen levels in a sample bottle change quickly due to the decomposition of organic material by microorganisms or the production of oxygen by algae and other plants in the sample. This will lower your DO reading.
ASSESSMENT. To serve as a usable method for manipulating oxygen in the lab, the system must reliably, precisely, and accurately produce the programmed DO patterns in the experiment tank. We tested the system by programming it to generate a diel oxygen cycle, with DO concentrations ranging from 3 mg L -1 to 7 mg L -1.
Oxygen is used by organisms in aerobic respiration, where energy is released by the combustion of sugar in the mitochondria. This form of oxygen can fit into the spaces between water molecules and is available to aquatic organisms. Fish, invertebrates, and other aquatic animals depend upon the oxygen dissolved in water.
Dissolved Oxygen and Temperature Lab Background: In this experiment, you will try to find out what happens to the amount of dissolved oxygen when the temperature of the water increases or decreases. This will help you understand how seasonal changes as well as long-term changes, such as climate change, might affect aquatic ecosystems.
Many lab activities can be conducted with our Wireless, PASPORT, or even ScienceWorkshop sensors and equipment. For assistance with substituting compatible instruments, contact PASCO Technical Support. We're here to help. Students will use a dissolved oxygen sensor to determine how different factors affect the concentration of dissolved oxygen ...